Accupyc II 1340
AccuPyc II TEC
Intergrated Thermo-Electric Temperature Control Solution
Materials tend to expand as they are heated, causing the same mass to occupy an increasing volume, thus lowering the substances density. Materials subjected to changing temperature may have a direct effect on volume, affecting accurate density determination.
The AccuPyc II TEC features a Peltier thermoelectric device for precise temperature control and stability. This instrument is an excellent option for use with temperature sensitive or viscous samples where environmental temperature cannot be adequately controlled.
Optional AccuPyc II TEC Software
Density Determination of Semi-Solid Bituminous Materials
This AccuPyc solution can be closely correlated (<0.15% difference) to results obtained with ASTM Test Method D70-09. The ASTM method is burdensome and time consuming. Our approach offers an expedited, more robust, operator-independent method, with results in minutes.
Average Density AccuPyc II TEC Solution (n=11) (g/cc) | %Relative Standard Deviation (n=11) | ASTM Method D70-9 Density (g/cc) | %Difference Between Methods | |
Sample A | 1.01906 | 0.03 | 1.01758 | 0.1453 |
Sample B | 1.02543 | 0.03 | 1.02536 | 0.0067 |
Sample C | 1.01821 | 0.07 | 1.01848 | 0.0263 |
Sample D | 1.02563 | 0.09 | 1.02576 | 0.0125 |
Samples varied in composition and viscosity
Height: 27.3 cm (10.7 in.) | Width: 28.7 cm (11.3 in) | Depth: 36.1 cm (14.2 in.) | Weight: 13.8 kg (30.5 lbs)
GSEM., JSC
Tel: +8424 62967539; Email: sales@gsemvn.com
Head Office: RM 630, 6th Floor, CIC Tower, 1 Nguyen Thi Due Str., Cau Giay Dist., Hanoi, Vietnam.
Registered Address: No. 18 Alley 313/6 Lane 313 Linh Nam, Hoang Mai, Hanoi